The National Institute for Quality (INACAL) has announced Draft Peruvian Metrological Standard PNMP 021:2026, «Non-invasive, Non-automated Sphygmomanometers,» consisting of: Part 1: Metrological and Technical Requirements, Part 2: Test Procedures, and Part 3: Test Report Format. The draft standard applies to non-invasive, non-automated sphygmomanometers, which are widely used in clinics, health centres, hospitals, community health campaigns, and primary healthcare services to monitor blood pressure. PNMP 021-1:2026 establishes stricter metrological and technical requirements, including: (i) a maximum permissible error of ±0.4 kPa (±3 mmHg); (ii) performance requirements under varying temperature conditions; (iii) tests for air leakage, hysteresis, pressure reduction rate, and stability; and (iv) requirements for mechanical safety and marking durability. PNMP 021-2:2026 sets out harmonized test procedures to support international traceability, including: 1. temperature testing; 2. testing under extreme storage conditions; 3. air leakage testing; 4. hysteresis testing; 5, durability testing; and 6. shock, vibration, and transport testing. PNMP 021-3:2026 establishes a standardized test report format for type approval and verification, supporting technical equivalence with the requirements of other members of the International Organization of Legal Metrology (OIML).